Format

Send to

Choose Destination
See comment in PubMed Commons below
Intensive Care Med. 2009 Aug;35(8):1368-76. doi: 10.1007/s00134-009-1467-7. Epub 2009 Apr 8.

A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators.

Author information

1
Medical Intensive Care Unit, AP-HP, Centre Hospitalier Albert Chenevier-Henri Mondor Hospital, 51 Avenue du Maréchal de Lattre de Tassigny, 94010 Creteil, France. arnaud.thille@hmn.aphp.fr

Abstract

OBJECTIVE:

To compare 13 commercially available, new-generation, intensive-care-unit (ICU) ventilators in terms of trigger function, pressurization capacity during pressure-support ventilation (PSV), accuracy of pressure measurements, and expiratory resistance.

DESIGN AND SETTING:

Bench study at a research laboratory in a university hospital.

METHODS:

Four turbine-based ventilators and nine conventional servo-valve compressed-gas ventilators were tested using a two-compartment lung model. Three levels of effort were simulated. Each ventilator was evaluated at four PSV levels (5, 10, 15, and 20 cm H2O), with and without positive end-expiratory pressure (5 cm H2O). Trigger function was assessed as the time from effort onset to detectable pressurization. Pressurization capacity was evaluated using the airway pressure-time product computed as the net area under the pressure-time curve over the first 0.3 s after inspiratory effort onset. Expiratory resistance was evaluated by measuring trapped volume in controlled ventilation.

RESULTS:

Significant differences were found across the ventilators, with a range of triggering delays from 42 to 88 ms for all conditions averaged (P < 0.001). Under difficult conditions, the triggering delay was longer than 100 ms and the pressurization was poor for five ventilators at PSV5 and three at PSV10, suggesting an inability to unload patient's effort. On average, turbine-based ventilators performed better than conventional ventilators, which showed no improvement compared to a bench comparison in 2000.

CONCLUSION:

Technical performance of trigger function, pressurization capacity, and expiratory resistance differs considerably across new-generation ICU ventilators. ICU ventilators seem to have reached a technical ceiling in recent years, and some ventilators still perform inadequately.

PMID:
19352622
PMCID:
PMC2873304
DOI:
10.1007/s00134-009-1467-7
[Indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Springer Icon for PubMed Central
    Loading ...
    Support Center