J Asthma. 2008 Dec;45(10):903-7. doi: 10.1080/02770900802386008.

Validation of automatic wheeze detection in patients with obstructed airways and in healthy subjects.

Author information

1
Baylor College of Medicine, Ben Taub General Hospital, 1504 Taub Loop, Houston, Texas 77030, USA. kkg@bcm.tmc.edu

Abstract

Computerized lung-sound analysis is a sensitive and quantitative method to identify wheezing by its typical pattern on spectral analysis. We evaluated the accuracy of the VRI, a multi-sensor, computer-based device with an automated technique of wheeze detection. The method was validated in 100 sound files from seven subjects with asthma or chronic obstructive pulmonary disease and seven healthy subjects by comparison of auscultation findings, examination of audio files, and computer detection of wheezes. Three blinded physicians identified 40 sound files with wheezes and 60 sound files without wheezes. Sensitivity and specificity were 83% and 85%, respectively. Negative predictive value and positive predictive value were 89% and 79%, respectively. Overall inter-rater agreement was 84%. False positive cases were found to contain sounds that simulate wheezes, such as background noises with high frequencies or strong noises from the throat that could be heard and identified without a stethoscope. The present findings demonstrate that the wheeze detection algorithm has good accuracy, sensitivity, specificity, negative predictive value and positive predictive value for wheeze detection in regional analyses with a single sensor and multiple sensors. Results are similar to those reported in the literature. The device is user-friendly, requires minimal patient effort, and, distinct from other devices, it provides a dynamic image of breath sound distribution with wheeze detection output in less than 1 minute.

PMID:
19085580
DOI:
10.1080/02770900802386008
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Taylor & Francis
Loading ...
Support Center