Talanta. 1997 Dec 19;45(2):331-41.

Derivative spectrophotometric determination of copper(II) in non-ionic micellar medium.

Author information

1
Department of Chemistry, University of Delhi, Delhi 110007, India.

Abstract

A photometric method for trace analysis of Cu(II) with 1-(2-pyridylazo)-2-naphthol (PAN) in presence of a neutral surfactant, Triton X-100, has been reported in normal and derivative modes. The molar absorption coefficient () and Sandell's sensitivity (S) of PAN-Cu(II) complex at 555 nm is 5.21 x 10(4) l mol(-1) cm(-1) and 1.22 ng cm(-2) respectively. The detection limit of Cu(II) is 4.0 ng ml(-1) and Beer's law is obeyed in the range 0.08-4.00 mug ml(-1) of the analyte. Copper content of a number of commercially available alcoholic beverages, biological and standard alloy samples is reported. The results obtained are in excellent agreement with those obtained using AAS.