Format

Send to

Choose Destination
See comment in PubMed Commons below
Appl Opt. 2008 Apr 20;47(12):2171-7.

Short coherence length produced by a spatial incoherent source applied for the Linnik-type interferometer.

Author information

1
Institute for Ultrafast Spectroscoy and Lasers, Department of Physics, City College and Graduate Center of the City University of New York, New York 10031, USA. zeylikov@sci.ccny.cuny.edu

Abstract

An axial resolution created by a spatial incoherent source is investigated theoretically and experimentally for the Linnik-type interferometer. The axial resolution in interference microscopy depends on both the temporal coherence length of the source and the objective numerical aperture (NA). Here the problem is treated in a more general situation by considering the spatial and temporal coherence of the illumination source which may be important for deep coherence imaging application. The results show that the axial resolution is degraded at the depth much less by using the optimal spectral bandwidth of the incoherent source and high-NA objectives.

PMID:
18425192
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Support Center