Format

Send to

Choose Destination
See comment in PubMed Commons below
Proteomics. 2008 Apr;8(8):1530-8. doi: 10.1002/pmic.200701146.

Precision enhancement of MALDI-TOF MS using high resolution peak detection and label-free alignment.

Author information

1
William and Mary Research Institute, College of William and Mary, Williamsburg, VA, USA.

Abstract

We have developed an automated procedure for aligning peaks in multiple TOF spectra that eliminates common timing errors and small variations in spectrometer output. Our method incorporates high-resolution peak detection, re-binning, and robust linear data fitting in the time domain. This procedure aligns label-free (uncalibrated) peaks to minimize the variation in each peak's location from one spectrum to the next, while maintaining a high number of degrees of freedom. We apply our method to replicate pooled-serum spectra from multiple laboratories and increase peak precision (t/sigma(t)) to values limited only by small random errors (with sigma(t) less than one time count in 89 out of 91 instances, 13 peaks in seven datasets). The resulting high precision allowed for an order of magnitude improvement in peak m/z reproducibility. We show that the CV for m/z is 0.01% (100 ppm) for 12 out of the 13 peaks that were observed in all datasets between 2995 and 9297 Da.

PMID:
18340636
PMCID:
PMC2413415
DOI:
10.1002/pmic.200701146
[Indexed for MEDLINE]
Free PMC Article