Measurement of the hemoglobin oxygen saturation level with spectroscopic spectral-domain optical coherence tomography.

Author information

Institute of Photonics and Optoelectronics, National Taiwan University, Taipei, Taiwan.


We report the measurement of the hemoglobin (Hb) oxygen saturation level in human blood with a spectroscopic spectral-domain optical coherence tomography (SSD-OCT) system based on the crossover behavior of Hb and oxyhemoglobin (HbO(2)) absorption coefficients around 800 nm. By calculating the ratio of the exponential decay constant of A-mode scan signal in the long-wavelength range (>800 nm) over that in the short-wavelength range (<800 nm), the relative oxygen saturation level of Hb can be calibrated. Such a relative level can be used for practical diagnosis application after a golden standard is built by comparing the variation of the OCT result with that of a conventional method, such as the use of a blood gas analyzer. In our experiment, the variation curve based on the SSD-OCT measurement, which has the advantages of noninvasive, real-time, and high-resolution measurements, is coincident with that of using a commercial blood gas analyzer.