J Biomed Opt. 2007 Nov-Dec;12(6):064003. doi: 10.1117/1.2821198.

Application of long-period-grating sensors to respiratory plethysmography.

Author information

1
Aston University, Photonics Research Group, Aston Triangle, Birmingham B4 7ET, United Kingdom. t.d.p.allsop@aston.ac.uk

Abstract

A series of in-line curvature sensors on a garment are used to monitor the thoracic and abdominal movements of a human during respiration. These results are used to obtain volumetric tidal changes of the human torso in agreement with a spirometer used simultaneously at the mouth. The curvature sensors are based on long-period gratings (LPGs) written in a progressive three-layered fiber to render the LPGs insensitive to the refractive index external to the fiber. A curvature sensor consists of the fiber long-period grating laid on a carbon fiber ribbon, which is then encapsulated in a low-temperature curing silicone rubber. The sensors have a spectral sensitivity to curvature, d lambda/dR from approximately 7-nm m to approximately 9-nm m. The interrogation technique is borrowed from derivative spectroscopy and monitors the changes in the transmission spectral profile of the LPG's attenuation band due to curvature. The multiplexing of the sensors is achieved by spectrally matching a series of distributed feedback (DFB) lasers to the LPGs. The versatility of this sensing garment is confirmed by it being used on six other human subjects covering a wide range of body mass indices. Just six fully functional sensors are required to obtain a volumetric error of around 6%.

PMID:
18163819
DOI:
10.1117/1.2821198
[Indexed for MEDLINE]