Format

Send to

Choose Destination
Magn Reson Med. 2008 Jan;59(1):221-7.

Single-shot dual-z-shimmed sensitivity-encoded spiral-in/out imaging for functional MRI with reduced susceptibility artifacts.

Author information

1
Brain Imaging and Analysis Center, Duke University Medical Center, Durham, North Carolina 27710, USA. truong@biac.duke.edu

Abstract

Blood oxygenation level-dependent (BOLD) functional MRI (fMRI) can be severely hampered by signal loss due to susceptibility-induced static magnetic field (B(0)) inhomogeneities near air/tissue interfaces. A single-shot spiral-in/out sequence with a z-shim gradient embedded between the two acquisitions was previously proposed to efficiently recover the signal. However, despite promising results, this technique had several limitations, which are addressed here as follows. First, by adding a second z-shim gradient before the spiral-in acquisition and optimizing both z-shim gradients slice-by-slice, a significantly more uniform signal recovery can be achieved. Second, by acquiring a B(0) map, the optimal z-shim gradients can be directly, efficiently, and accurately determined for each subject. Third, by complementing the z-shimming approach with sensitivity encoding (SENSE), the in-plane spatial resolution can be increased and, hence, susceptibility artifacts further reduced, while maintaining a high temporal resolution for fMRI applications. These advantages are demonstrated in human functional studies.

PMID:
18050341
DOI:
10.1002/mrm.21473
[Indexed for MEDLINE]
Free full text

Supplemental Content

Full text links

Icon for Wiley
Loading ...
Support Center