Format

Send to

Choose Destination
J Proteome Res. 2007 Nov;6(11):4200-9. Epub 2007 Sep 29.

iTRAQ reagent-based quantitative proteomic analysis on a linear ion trap mass spectrometer.

Author information

1
Department of Biochemistry, Molecular Biology, and Biophysics, Center for Mass Spectrometry and Proteomics, University of Minnesota, Minneapolis, Minnesota 55455, USA. tgriffin@umn.edu

Abstract

For proteomic analysis using tandem mass spectrometry, linear ion trap instruments provide unsurpassed sensitivity but unreliably detect low mass peptide fragments, precluding their use with iTRAQ reagent-labeled samples. Although the popular LTQ linear ion trap supports analyzing iTRAQ reagent-labeled peptides via pulsed Q dissociation, PQD, its effectiveness remains questionable. Using a standard mixture, we found careful tuning of relative collision energy necessary for fragmenting iTRAQ reagent-labeled peptides, and increasing microscan acquisition and repeat count improves quantification but identifies somewhat fewer peptides. We developed software to calculate abundance ratios via summing reporter ion intensities across spectra matching to each protein, thereby providing maximized accuracy. Testing found that results closely corresponded between analysis using optimized LTQ-PQD settings plus our software and using a Qstar instrument. Thus, we demonstrate the effectiveness of LTQ-PQD analyzing iTRAQ reagent-labeled peptides, and provide guidelines for successful quantitative proteomic studies.

PMID:
17902639
PMCID:
PMC2533114
DOI:
10.1021/pr070291b
[Indexed for MEDLINE]
Free PMC Article

Supplemental Content

Full text links

Icon for American Chemical Society Icon for PubMed Central
Loading ...
Support Center