High-performance liquid chromatographic determination of rifapentine and its metabolite in human plasma by direct injection into a shielded hydrophobic phase column

J Chromatogr. 1991 Aug 16;553(1-2):35-40. doi: 10.1016/s0021-9673(01)88469-7.

Abstract

A simple high-performance liquid chromatographic method for determination of rifapentine, a cyclopentyl semisynthetic analogue of rifamycin belonging to the class of piperazinyl hydrazone derivatives of 3-formylrifamycin SV, and its metabolite, 25-desacetylrifapentine, in human plasma was developed using direct injection of the sample onto a Supelco LC HISEP column. The mean recovery was 100.3% for rifapentine and 99.7% for the metabolite and the precision of the assays was 3% and 7%, respectively. The limit of determination was 0.2 micrograms/ml and the method was validated for concentrations up to 64 micrograms/ml for rifapentine and 32 micrograms/ml for the metabolite. The results correlated well with those of the microbiological assay with Sarcina lutea as test organism.

MeSH terms

  • Anti-Bacterial Agents / blood*
  • Anti-Bacterial Agents / metabolism
  • Chromatography, High Pressure Liquid / instrumentation
  • Humans
  • Reproducibility of Results
  • Rifampin / analogs & derivatives*
  • Rifampin / blood
  • Rifampin / metabolism
  • Rifampin / pharmacology
  • Sarcina / drug effects
  • Spectrophotometry, Ultraviolet

Substances

  • Anti-Bacterial Agents
  • 25-desacetylrifapentine
  • Rifampin
  • rifapentine