Clin Vaccine Immunol. 2007 Aug;14(8):1050-2. Epub 2007 Jun 13.

Clinical evaluation of the SD Bioline influenza virus antigen test for rapid detection of influenza viruses A and B in children and adults during the influenza season.

Author information

1
Department of Pediatrics, Korea University, Seoul 136-705, Korea.

Abstract

The performance of the SD Bioline rapid antigen test kit for influenza virus detection was evaluated with 295 respiratory specimens during the influenza season. The overall sensitivity and specificity of the SD Bioline test were 61.9% and 96.8% for the influenza A virus antigen and 54.5% and 100% for the influenza B virus antigen, respectively. The results were consistent with peak influenza activities.

PMID:
17567765
PMCID:
PMC2044486
DOI:
10.1128/CVI.00465-06