Otolaryngol Head Neck Surg. 2007 Jun;136(6):966-71.

Comparing pure-tone audiometry and auditory steady state response for the measurement of hearing loss.

Author information

1
Department of Otolaryngology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.

Abstract

OBJECTIVE:

To compare pure-tone audiometry and auditory steady state response (ASSR) to measure hearing loss based on the severity of hearing loss in frequencies.

STUDY DESIGN AND SETTING:

A total of 105 subjects (168 ears, 64 male and 41 female) were enrolled in this study. We determined hearing level by measurement of pure-tone audiometry and ASSR on the same day for each subject.

RESULTS:

Pure-tone audiometry and ASSR were highly correlated (r=0.96). The relationship is described by the equation PTA=1.05 x mean ASSR - 7.6. When analyzed according to the frequencies, the correlation coefficients were 0.94, 0.95, 0.94, and 0.92 for 0.5, 1, 2, and 4 kHz, respectively.

CONCLUSION:

From this study, authors could conclude that pure-tone audiometry and ASSR showed very similar results and indicated that ASSR may be a good alternative method for the measurement of hearing level in infants and children, for whom pure-tone audiometry is not appropriate.