Format

Send to

Choose Destination
Biomaterials. 2007 Sep;28(25):3594-607. Epub 2007 Apr 5.

Neural probe design for reduced tissue encapsulation in CNS.

Author information

1
Department of Biomedical Engineering, University of Michigan, 2212 Lurie Biomedical Engineering Building, 1101 Beal Avenue, Ann Arbor, MI 48109-2099, USA.

Abstract

This study investigated relationships between a microscale neural probe's size and shape and its chronic reactive tissue response. Parylene-based probes were microfabricated with a thick shank (48 microm by 68 microm) and an integrated thin lateral platform (5 microm by 100 microm, either solid or one of three lattice sizes). Devices were implanted in rat cerebral cortex for 4 weeks before immunostaining for neurons, astrocytes, microglia, fibronectin, laminin, and neurofilament. While nonneuronal density (NND) generally increased and neuronal density decreased within 75 microm of a probe interface compared to unimplanted control regions, there were significant differential tissue responses within 25 microm of the platform's lateral edge compared to the shank. The NND in this region of the lateral edge was less than one-third of the corresponding region of the shank (129% and 425% increase, respectively). Moreover, neuronal density around the platform lateral edge was about one-third higher than at the shank (0.70 and 0.52 relative to control, respectively). Also, microglia reactivity and extracellular protein deposition was reduced at the lateral edge. There were no significant differences among platform designs. These results suggest that neural probe geometry is an important parameter for reducing chronic tissue encapsulation.

[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Elsevier Science
Loading ...
Support Center