Format

Send to

Choose Destination
See comment in PubMed Commons below
Opt Lett. 2007 Mar 15;32(6):632-4.

Concept of a high-resolution miniature spectrometer using an integrated filter array.

Author information

1
National Laboratory for Infrared Physics, Shanghai Institute of Technical Physics, Chinese Academy of Sciences. Wangshw@mail.sitp.ac.cn

Abstract

A high-resolution miniature spectrometer has been demonstrated by utilizing a 128-channel integrated filter array, fabricated by using the combinatorial deposition technique, as a dispersive component whose passbands range from 722.0 to 880.0 nm with a bandwidth (or spectral resolution) from 1.7 to 3.8 nm and an average channel interval of 1.2 nm. The miniature spectrometer is smaller than 1 cm3 without any moving parts. This kind of miniature spectrometer has the advantages of very low payload, high resolution, and high reliability simultaneously, which are especially urgently needed for space applications.

PMID:
17308584
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Support Center