Magn Reson Med. 2007 Feb;57(2):331-7.

Diffusion imaging of the in vivo heart using spin echoes--considerations on bulk motion sensitivity.

Author information

  • 1Institute for Biomedical Engineering, University of Zurich and Swiss Federal Institute of Technology, Zurich, Switzerland.

Abstract

Cardiac diffusion MRI based on stimulated-echo acquisition mode (STEAM) techniques is hampered by its inherent low signal-to-noise ratio (SNR) efficiency. Diffusion imaging using standard spin-echo (SE) techniques, on the other hand, offers higher SNRs but has been considered impractical for the beating heart due to excessive signal attenuation from cardiac bulk motion. In this work the effect of systolic cardiac motion on different diffusion-encoding schemes was studied in detail. Numerical simulations based on in vivo motion data (acquired by MRI tagging techniques) demonstrate an up to 10-fold decrease in bulk motion sensitivity of the diffusion encoding if the first-order moment of the diffusion-encoding gradients is nullified. It is shown that the remaining systolic phase pattern on the myocardium does not influence the magnitude images if the spatial resolution is chosen to be higher than 4 mm. Given these relatively low resolution requirements, we obtained in vivo diffusion-weighted (DW) short-axis images from four healthy volunteers using an SE-based diffusion-encoding sequence with excitation and refocusing in orthogonal planes for field of view (FOV) reduction. The results showed no significant signal loss due to cardiac motion, and the direction of the principal eigenvalues was found to be in good agreement with known myocardial fiber orientation.

Copyright (c) 2007 Wiley-Liss, Inc.

[PubMed - indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Wiley
    Loading ...
    Write to the Help Desk