Format

Send to

Choose Destination
See comment in PubMed Commons below
Kidney Int. 2007 Jan;71(1):74-8. Epub 2006 Nov 1.

Plasma creatinine determination in mice and rats: an enzymatic method compares favorably with a high-performance liquid chromatography assay.

Author information

1
Medical Research Center, University of Heidelberg, Theodor-Kutzer-Ufer, Mannheim, Germany.

Abstract

The use of the colorimetric Jaffé method for the measurement of creatinine in mouse and rat plasma has been criticized as prior studies have shown a dramatic overestimation. We compared a colorimetric picric acid, an enzymatic, and a high-performance liquid chromatography (HPLC) method to assess their appropriateness for routine measurements of creatinine in plasma of healthy and diseased mice (n=61) and rats (n=56). For the colorimetric Jaffé method a pronounced overestimation is confirmed. Additionally the method showed interference with hemoglobin already in a very low, non-visible concentration range in rat plasma. The enzymatic measurement demonstrated a hemoglobin interference in mice, only when hemolysis was visible. The comparison between HPLC and the enzymatic measurement gave a good agreement between both methods in both species. Therefore the enzymatic method fulfills the requirements for a routine screening test for plasma creatinine in healthy as well as diseased mice and rats Kiover a broad concentration range.

PMID:
17082757
DOI:
10.1038/sj.ki.5001988
[Indexed for MEDLINE]