Magn Reson Med. 2006 Nov;56(5):1019-32.

Real-time rigid body motion correction and shimming using cloverleaf navigators.

Author information

1
Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts 02129, USA. andre@nmr.mgh.harvard.edu

Abstract

Subject motion during scanning can greatly reduce MRI image quality and is a major reason for discarding data in both clinical and research scanning. The quality of the high-resolution structural data used for morphometric analysis is especially compromised by subject movement because high-resolution scans are of longer duration. A method is presented that measures and corrects rigid body motion and associated first-order shim changes in real time, using a pulse sequence with embedded cloverleaf navigators and a feedback control mechanism. The procedure requires a 12-s preliminary mapping scan. A single-path, 4.2-ms cloverleaf navigator is inserted every repetition time (TR) after the readout of a 3D fast low-angle shot (FLASH) sequence, requiring no additional RF pulses and minimally impacting scan duration. Every TR, a rigid body motion estimate is made and a correction is fed back to adjust the gradients and shim offsets. Images are corrected and reconstructed on the scanner computer for immediate access. Correction for between-scan motion can be accomplished by using the same reference map for each scan repetition. Human and phantom tests demonstrated a consistent improvement in image quality if motion occurred during the acquisition.

PMID:
17029223
DOI:
10.1002/mrm.21038
[Indexed for MEDLINE]
Free full text

Supplemental Content

Full text links

Icon for Wiley
Loading ...
Support Center