IEEE Trans Med Imaging. 2006 Oct;25(10):1341-7.

Ultrahigh-field MRI whole-slice and localized RF field excitations using the same RF transmit array.

Author information

1
Department of Radiology, University of Pittsburgh, Pittsburgh, PA 15213, USA. tsi2@pitt.edu

Abstract

In this paper, a multiport driving mechanism is numerically implemented at ultra high-field (UHF) magnetic resonance imaging (MRI) to provide 1) homogenous whole-slice (axial, sagittal, or coronal) and 2) highly localized radio frequency (RF) field excitation within the same slices, all with the same RF transmit array (here chosen to be a standard transverse electromagnetic (TEM) resonator/coil). The method is numerically tested using a full-wave model of a TEM coil loaded with a high-resolution/18-tissue/anatomically detailed human head mesh. The proposed approach is solely based on electromagnetic and phased array antenna theories. The results demonstrate that both homogenous whole-slice as well as localized RF excitation can be achieved within any slice of the head at 7 T (298 MHz for proton imaging).

PMID:
17024837
[Indexed for MEDLINE]