Cough. 2006 Sep 27;2:7.

Evaluation of a new self-contained, ambulatory, objective cough monitor.

Author information

1
Department of Pediatrics, Penn State College of Medicine, Hershey, PA, USA. ipaul@psu.edu

Abstract

OBJECTIVE AND BACKGROUND:

Objective monitoring of cough may be preferred to subjective reporting of the symptom in clinical and research settings. Therefore, a self-contained, ambulatory cough monitoring system is needed that is non-invasive, usable for children and adults of all ages, inexpensive, and highly accurate with easy to use analysis software.

METHODOLOGY:

After development of a new device, 15 subjects with frequent coughing were recorded with the novel cough monitor and a simultaneous video recording in order to validate the monitor compared with a gold standard. Two investigators independently analyzed the recordings and counted the number of coughs during the study period from both the cough monitor and the video recording.

RESULTS:

When measuring agreement between the two investigators, the sample concordance correlation coefficient for audio counts was 0.998 (p < 0.001). In the comparison of video counts, the sample concordance correlation coefficient was 0.997 (p < 0.001). For the comparison of investigator 1's video counts to the corresponding audio counts, the sample concordance correlation coefficient was 0.968 (p = 0.026). For the comparison of investigator 2's video counts to the corresponding counts, the sample concordance correlation coefficient was 0.973 (p = 0.015).

CONCLUSION:

We have developed and piloted a new, valid, and reproducible method of objectively recording and analyzing cough. This device appears to be useful for subjects of any age and in clinical and research settings.