J Bone Joint Surg Br. 2006 Aug;88(8):1003-5.

Serum aluminium and cobalt levels after ceramic-on-ceramic and metal-on-metal total hip replacement.

Author information

1
Department of Orthopaedics, Medical University of Vienna Austria. alexander.gruebl@meduniwien.ac.at

Abstract

In a randomised study, 28 patients with a mean age of 62.2 years (32 to 81) with osteoarthritis or avascular necrosis of the hip received either a ceramic-on-ceramic or a metal-on-metal total hip replacement. Apart from the liners the acetabular and femoral components were made of Ti-Al-Nb alloy. The serum aluminium and cobalt levels were measured before, and at one year after surgery. The 15 patients in the ceramic-on-ceramic group had a median pre-operative aluminium level of 1.3 microg/l (0.25 to 8.4) and a cobalt level below the detection limit. At one year the aluminium level was 1.1 microg/l (0.25 to 2.3) and the cobalt level was 0.4 microg/l (0.15 to 0.7). The 13 patients in the metal-on-metal group had a median pre-operative aluminium level of 1.9 microg/l (0.25 to 4.4) and a cobalt level below the detection limit. At one year the median aluminium level was 0.9 microg/l (0.25 to 3.9) whereas the cobalt level was 1.4 microg/l (0.5 to 10.5). This increase in the cobalt level at one year was significant (p < 0.001). Our findings indicate that ceramic-on-ceramic bearings do not cause elevated levels of serum aluminium in the first post-operative year.

PMID:
16877596
DOI:
10.1302/0301-620X.88B8.17870
[Indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for HighWire
    Loading ...
    Support Center