Gamma-ray imaging with a Si/CsI(Tl) Compton detector

Appl Radiat Isot. 2006 Dec;64(12):1648-54. doi: 10.1016/j.apradiso.2006.05.015. Epub 2006 Jul 11.

Abstract

We present results from Compton imaging of gamma-ray sources using an instrument constructed from thin silicon scattering detectors and CsI(Tl) absorbing detectors. We have successfully imaged single and double point sources for several common radioactive isotopes ((137)Cs, (60)Co, (22)Na, (54)Mn). The measured angular resolution is 11.6( composite function) FWHM at 662keV. In parallel with the hardware effort, a GEANT4-based simulation code was developed. Comparisons between real and simulated data are discussed.

MeSH terms

  • Cesium*
  • Equipment Design
  • Gamma Rays*
  • Iodides*
  • Radiation Monitoring / instrumentation
  • Silicon*
  • Spectrometry, Gamma / instrumentation*

Substances

  • Iodides
  • Cesium
  • cesium iodide
  • Silicon