Efficiency-corrected dose verification with thermoluminescence dosemeters in heavy-ion beams

Radiat Prot Dosimetry. 2006;120(1-4):361-4. doi: 10.1093/rpd/nci581. Epub 2006 May 18.

Abstract

One of the most essential difficulties in heavy-ion dosimetry by means of thermoluminescence dosemeters (TLDs)--often seen as a serious disadvantage of TLD utilisation--regards the changing TL-efficiency with increasing linear energy transfer (LET) of the particle. This behaviour leads to a falsification of absorbed dose that can be significant for many applications, e.g. in space or radiotherapeutic dosimetry. The high-temperature TL emission of LiF:Mg,Ti TL detectors can be exploited to obtain information about the LET of the heavy-ion radiation field under study. The high-temperature ratio (HTR) is used as a parameter to determine average LET. To correct the absorbed dose according to the TL-efficiency, the detailed dependence of HTR- and TL-efficiency on LET was recorded. These investigations were accomplished at the Heavy Ion Medical Accelerator (HIMAC) in Chiba, Japan, with a variety of high-energy ion beams (helium, carbon, neon, silicon and iron) ranging in LET from 2.2 to 393 keV microm(-1). The obtained relationships HTR vs. LET and TL-efficiency vs. LET were combined into a TL efficieny vs. HTR relationship. This enables correction of the absorbed dose (HTR-B method). The methodology is demonstrated by means of TLD 700 ((7)LiF:Mg,Ti) measurements in carbon beams of 290 and 400 MeV n(-1) available from HIMAC.

MeSH terms

  • Algorithms*
  • Artifacts*
  • Dose-Response Relationship, Radiation
  • Equipment Design
  • Equipment Failure Analysis
  • Heavy Ions*
  • Materials Testing
  • Radiation Dosage
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thermoluminescent Dosimetry / instrumentation*
  • Thermoluminescent Dosimetry / methods*