Phys Med Biol. 2006 May 21;51(10):N205-10. Epub 2006 Apr 26.

Evaluation of a new electromagnetic tracking system using a standardized assessment protocol.

Author information

1
Center of Biomedical Engineering and Physics, Medical University of Vienna, Austria. Ludwig-Boltzmann Institute of Nuclear Medicine, Vienna, Austria.

Abstract

This note uses a published protocol to evaluate a newly released 6 degrees of freedom electromagnetic tracking system (Aurora, Northern Digital Inc.). A practice for performance monitoring over time is also proposed. The protocol uses a machined base plate to measure relative error in position and orientation as well as the influence of metallic objects in the operating volume. Positional jitter (E(RMS)) was found to be 0.17 mm +/- 0.19 mm. A relative positional error of 0.25 mm +/- 0.22 mm at 50 mm offsets and 0.97 mm +/- 1.01 mm at 300 mm offsets was found. The mean of the relative rotation error was found to be 0.20 degrees +/- 0.14 degrees with respect to the axial and 0.91 degrees +/- 0.68 degrees for the longitudinal rotation. The most significant distortion caused by metallic objects is caused by 400-series stainless steel. A 9.4 mm maximum error occurred when the rod was closest to the emitter, 10 mm away. The improvement compared to older generations of the Aurora with respect to accuracy is substantial.

PMID:
16675856
DOI:
10.1088/0031-9155/51/10/N01
[Indexed for MEDLINE]