J Chromatogr. 1992 Feb 28;593(1-2):21-4.

Automated high-performance liquid chromatographic method for the determination of rifampicin in plasma.

Author information

1
Department of Pharmacology, Faculty of Medicine, UOFS, Bloemfontein, South Africa.

Abstract

Due to the unstable nature of rifampicin, a rapid automated high-performance liquid chromatographic method had to be developed for the analysis of a large number of plasma samples generated during a bioavailability trial. Extraction and injection of the samples were automatically done by a sample preparation system using C2, 100 mg Bond Elut extraction columns. The extracts were chromatographed on a 4-microns reversed-phase C18 column with a citrate buffer and acetonitrile as mobile phase. The analytes were detected at 342 nm. Calibration curves were linear to at least 20 micrograms/ml and the limit of quantification was 0.16 micrograms/ml.

PMID:
1639906
DOI:
10.1016/0021-9673(92)80260-2
[Indexed for MEDLINE]