J Refract Surg. 2005 Sep-Oct;21(5 Suppl):S655-7.

Incidence of complications during flap creation in LASIK using the NIDEK MK-2000 microkeratome in 26,600 cases.

Author information

1
Codet Aris Vision Institute, Toluca, Mexico. carlos.carrillo@arisvision.com.mx

Abstract

PURPOSE:

To evaluate the incidence of intraoperative complications using the NIDEK MK-2000 microkeratome during flap creation in LASIK.

METHODS:

The incidence of intraoperative flap complications during LASIK using the NIDEK MK-2000 microkeratome was retrospectively studied in 26,600 procedures performed in 5 outpatient excimer laser surgery centers.

RESULTS:

A total of 65 (0.244%) complications were identified: 23 (0.086%) eyes had free caps, 13 (0.049%) eyes had an incomplete pass, 13 (0.049%) eyes had an epithelial defect, 11 (0.041%) eyes had buttonhole, and 5 (0.019%) eyes had irregular flap. The remaining 26,535 (99.756%) eyes had uneventful flap creation.

CONCLUSIONS:

Use of the NIDEK MK-2000 microkeratome resulted in a low incidence of intraoperative complications, making it a safe and reliable device for creating flaps during LASIK.