Send to

Choose Destination
Opt Lett. 2005 Oct 1;30(19):2647-9.

Improving the spectral resolution of wedged etalons and linear variable filters with incidence angle.

Author information

Department of Electrical and Computer Engineering, University of Colorado, Campus Box 425, Boulder, Colorado 80309, USA.


Compact spectrometers fabricated by placing a wedged etalon or a linear variable filter in front of a linear photodiode array typically do not exhibit the spectral resolution of a parallel filter of the same finesse. A theoretical analysis of the beam emerging from a wedged etalon is shown to predict a comalike aberration, explaining the loss in resolution. This analysis also predicts a focused beam waist whose axial position depends on the incidence angle, indicating that resolution can be improved by placing this waist on the detector. These results are experimentally confirmed in a multicavity linear variable filter to achieve 0.09 nm spectral resolution at 1550 nm.