Depth–dose distributions for a spread-out Bragg peak (SOBP, red), its constituent pristine Bragg peaks (blue), and a 10 MV photon beam (black). The SOBP dose distribution is created by adding the contributions of individually modulated pristine Bragg peaks. The penetration depth, or range, measured as the depth of the distal 90% of plateau dose, of the SOBP dose distribution is determined by the range of the most distal pristine peak (labeled ‘Pristine peak'). The modulation width, measured as the distance between the proximal and distal 90% of plateau dose values, of the SOBP dose distribution is controlled by varying the number and intensity of pristine Bragg peaks that are added, relative to the most distal pristine peak, to form the SOBP. The dashed lines (black) indicate the clinical acceptable variation in the plateau dose of ±2%. The dot–dashed lines (green) indicate the 90% dose and spatial, range and modulation width, intervals. The SOBP dose distribution of even a single field can provide complete target volume coverage in depth and lateral dimensions, in sharp contrast to a single photon dose distribution; only a composite set of photon fields can deliver a clinical target dose distribution. Note the absence of dose beyond the distal fall-off edge of the SOBP.