Appl Opt. 2005 Jul 10;44(20):4228-38.

Wave-front retrieval from Hartmann test data.

Author information

  • 1Centro de Investigaciones en Optica, A. C., Apdo. Postal 1-948, 37000 Leon, Gto., México.


In the classical Hartmann test the wave front is obtained by integration of the transverse aberrations, joining the sampled points by small straight segments, in the so-called Newton integration. This integration is performed along straight lines joining the holes on the Hartmann screen. We propose a modification of this procedure, considering the cells of four holes of the Hartmann screen to fit a small second-power wave front recovering each square. This procedure has some important advantages, as described here.