Am J Phys Med Rehabil. 2005 Aug;84(8):563-75.

A clinical comparison of variable-damping and mechanically passive prosthetic knee devices.

Author information

1
Department of Physical Medicine and Rehabilitation, Harvard Medical School, Spaulding Rehabilitation Hospital, Boston, Massachusetts, USA.

Abstract

OBJECTIVE:

Although variable-damping knee prostheses offer some improvements over mechanically passive prostheses to transfemoral amputees, there is insufficient evidence that such prostheses provide advantages at self-selected walking speeds. In this investigation, we address this question by comparing two variable-damping knees, the hydraulic-based Otto Bock C-leg and the magnetorheological-based Ossur Rheo, with the mechanically passive, hydraulic-based Mauch SNS.

DESIGN:

For each prosthesis, metabolic data were collected on eight unilateral amputees walking at self-selected speeds across an indoor track. Furthermore, kinetic, kinematic, and electromyographic data were collected while walking at self-selected speeds across a 10-m walkway in a laboratory.

RESULTS:

When using the Rheo, metabolic rate decreases by 5% compared with the Mauch and by 3% compared with the C-leg. Furthermore, for the C-leg and Rheo knee devices, we observe biomechanical advantages over the mechanically passive Mauch. These advantages include an enhanced smoothness of gait, a decrease in hip work production, a lower peak hip flexion moment at terminal stance, and a reduction in peak hip power generation at toe-off.

CONCLUSION:

The results of this study indicate that variable-damping knee prostheses offer advantages over mechanically passive designs for unilateral transfemoral amputees walking at self-selected ambulatory speeds, and the results further suggest that a magnetorheological-based system may have advantages over hydraulic-based designs.

[Indexed for MEDLINE]