Purpose: To search for individualized scan protocols that provide adequate diagnostic information with minimal radiation exposure for abdominal CT in adults.

Material and methods: Beginning with standard settings (120 kVp, 200 mA, 0.75 s, CTDI (w = 11.7 mGy)), which are in the lowest quartile of the actual German radiation dose survey, 119 adults were examined using a single-row detector helical CT with 8 mm collimation and a pitch of 1.5. The individualized scan protocol was defined by repeating the reference scan up to 3 times with different radiation doses following a predefined iteration scheme. The image qualities with standard dose and with individualized dose were assessed independently by 3 radiologists blinded to the exposure parameters. The individualized dose was correlated with the diameter of the patients.

Results: The patient diameters varied from 16 to 35 cm and correlated with individualized radiation doses CTDI (w) from 6.4 to 17.6 mGy. We found an exponential correlation (y = 52.3 x e (0.05 x); r(2) = 0.48; p < 0.001) between required tube current (mA) and a. p. diameter (cm) in the epigastric region in adults.

Conclusion: Maintaining constant exposure parameters results in unnecessary radiation exposure in patients with a diameter of less than 27 cm measured in anterior-posterior direction in the epigastric area. The radiation exposure in adult abdominal CT can be reduced up to 45 % by individualized selection of the tube current without adversely affecting diagnostic performance.

Publication types

  • Clinical Trial
  • English Abstract

MeSH terms

  • Adult
  • Dose-Response Relationship, Radiation
  • Female
  • Humans
  • Male
  • Quality Assurance, Health Care / methods*
  • Radiation Dosage*
  • Radiation Injuries / etiology
  • Radiation Injuries / prevention & control
  • Radiation Protection / methods*
  • Radiographic Image Enhancement / methods*
  • Radiography, Abdominal / methods*
  • Radiometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, Spiral Computed / adverse effects
  • Tomography, Spiral Computed / methods*