Comparison of bipolar and unipolar radiofrequency ablation in an in vivo experimental model

Eur J Cardiothorac Surg. 2005 Jul;28(1):76-80; discussion 80-2. doi: 10.1016/j.ejcts.2005.02.028. Epub 2005 Apr 7.

Abstract

Objective: Linear atrial radiofrequency lesions have been used effectively for the treatment of atrial fibrillation. In most cases an endocardial approach has been suggested. A method for epicardial placement of lesions would reduce the complexity of these procedures. We compared lesions created in ovine hearts in vivo using irrigated bipolar or unipolar handheld radiofrequency ablation devices.

Methods: Radiofrequency lesions were produced around a left pulmonary vein, around the left atrial appendage and in the free wall of the right ventricle in ovine hearts. All lesions were created in the beating heart. A bipolar clamping device (n = 7) or a handheld unipolar device (n = 6) was used. Measurements of local electrograms and pacing thresholds were performed before and after ablation at each site to assess the electrical integrity of lesions. Tetrazolium and digital image analysis were used to assess lesion geometry.

Results: In atrial tissue continuous transmural lesions were achieved more often with the bipolar than with the unipolar device (92.3 vs. 33.3%, P < 0.02). In atrial tissue the reduction in signal amplitude caused by the lesions was significantly larger with the bipolar than the unipolar device (87.6+/-9.4% vs. 60.6+/-23.7% reduction, P < 0.01). There was a significant relationship between loss of pacing capture and lesion transmurality (P < 0.05). The bipolar device created narrower lesions than the unipolar device (4.1+/-0.9 mm vs. 5.9+/-2.1 mm, P < 0.001).

Conclusions: The bipolar clamping device produces narrower lesions which are more likely to be transmural and lead to electrical isolation of ablated tissue than those produced by the unipolar device. However, both devices failed to consistently produce transmural lesions using the epicardial beating heart technique studied, particularly in thicker tissues. High output pacing within the ablated tissue partially predicts lesion transmurality and be a guide to the need for further ablation. However, endocardial ablation or transmural bipolar ablation are likely to remain the techniques of choice for linear radiofrequency ablation in the atria until improved techniques are developed.

Publication types

  • Comparative Study

MeSH terms

  • Animals
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Catheter Ablation / instrumentation
  • Catheter Ablation / methods*
  • Disease Models, Animal
  • Electrocardiography
  • Heart Atria / surgery
  • Heart Ventricles / pathology
  • Heart Ventricles / surgery
  • Male
  • Pericardium / pathology
  • Pericardium / surgery
  • Pulmonary Veins / physiopathology
  • Sheep