Optimal stimulus intensity and reliability of air stimulation technique for elicitation of laryngo-upper esophageal sphincter contractile reflex

Ann Otol Rhinol Laryngol. 2005 Mar;114(3):223-8. doi: 10.1177/000348940511400310.

Abstract

To determine the optimal air stimulus intensity and duration for elicitation of the laryngo-upper esophageal sphincter (UES) contractile reflex, we studied 37 healthy volunteers 20 to 81 years of age. A sleeve device monitored the UES pressure. For laryngeal stimulation, we used an air stimulator unit (Pentax AP-4000) that incorporated a nasolaryngeal endoscope. The arytenoids and interarytenoid areas were stimulated at least three times by three different stimuli: 6-mm Hg air pulse with 50-ms duration, 10-mm Hg air pulse with 50-ms duration, and 6-mm Hg air pulse with 2-second duration. Of 1,165 air stimulations, 1,041 resulted in mucosal deflections. Of these, 451 resulted in an abrupt increase in UES pressure. The response/deflection ratio for 6-mm Hg stimulation with 2-second duration was significantly higher than those for air pulses with 50-ms duration (p < .001). We conclude that although the laryngo-UES contractile reflex can be elicited by an air pulse with 50-ms duration, this ultrashort stimulation is not reliable. Using longer-duration pulses (at least 2 seconds) improves the reliability of elicitation of the laryngo-UES contractile reflex.