J Clin Virol. 2005 May;33(1):19-24.

Clinical evaluation of real-time PCR assays for rapid diagnosis of SARS coronavirus during outbreak and post-epidemic periods.

Author information

1
Department of Microbiology, The University of Hong Kong, University Pathology Building, Queen Mary Hospital Compound, Pokfulam, Hong Kong SAR, PR China. wcyam@hku.hk

Abstract

BACKGROUND:

The protocols of WHO network laboratories facilitated development of rapid diagnosis for SARS coronavirus (CoV) using reverse transcription (RT)-PCR assays. However, several reports have shown that conventional and real-time PCR assays were very specific for SARS CoV but lack sensitivity depending on the assay, specimen, and time course of disease.

OBJECTIVE:

To evaluate an automatic nucleic acid extraction system and two standardized real-time PCR assays for rapid diagnosis of SARS CoV during outbreak and post-epidemic periods in Hong Kong.

STUDY DESIGN:

Specimens from clinically suspected SARS patients collected during outbreak and post-epidemic periods were tested by an automatic nucleic acid extraction system followed by our first generation conventional RT-PCR and two standardized real-time PCR assays (Artus GmbH, Germany and Roche Diagnostics, Germany). Paired serum samples were assayed for increasing titer against SARS CoV.

RESULTS:

In the SARS epidemic, Artus and Roche PCR assays exhibited sensitivities of 87% and 85% for respiratory specimens (n = 64), 91% and 88% for stool (n = 44), and 82% for urine (n = 29). A specificity of 100% was exhibited by both PCR assays except Artus attained only a 92% specificity for stool. For post-epidemic period, no SARS CoV was identified among 56 respiratory specimens by all PCR assays. Inhibitors to PCR assays were detected at an average rate of 7-8% among 202 clinical specimens.

CONCLUSION:

This study highlights the high throughput and performance of automatic RNA extraction in coordination with standardized real-time PCR assays suitable for large-scale routine diagnosis in case of future SARS epidemic. As no SARS CoV was detected among specimens collected during post-epidemic period, the positive predictive value of real-time PCR assays for detection of SARS CoV during low epidemic requires further evaluation.

PMID:
15797361
DOI:
10.1016/j.jcv.2004.09.029
[Indexed for MEDLINE]