J Cardiovasc Magn Reson. 2004;6(4):885-94.

Dynamic real-time architecture in magnetic resonance coronary angiography--a prospective clinical trial.

Author information

1
Division of Cardiovascular Medicine, Department of Medicine, Stanford University Medical Center, Stanford, California 94305-5233, USA. pyang@cvmed.stanford.edu

Abstract

OBJECTIVES:

A dynamic real-time (dRT) architecture has been developed to address limitations in magnetic resonance coronary angiography (MRCA). A prospective clinical trial of 45 patients suspected of coronary artery disease was conducted to determine clinical utility of this integrated real-time system.

BACKGROUND:

Clinical implementation of MRCA is not performed routinely today. However, improved anatomic coverage, image quality, and scan flexibility may enhance its clinical utility. A novel real-time architecture addresses these challenges through instantaneous reconfiguration between real-time (RT) and high-resolution (HR) imaging sequences with dynamic selection of the desired element on a custom-designed receiver coil.

METHODS:

A total of 45 subjects were recruited consecutively to evaluate scan time, anatomic coverage, image quality, and detection of coronary lesions. Using a modern PC, the dRT switches from RT to gated HR imaging sequence in one repetition time (39 ms). Magnetic resonance imaging (MRI) scanning was performed using a custom-designed coronary coil consisting of two four-inch phase-array circular elements enabled with real-time selection of the desired coil element.

RESULTS:

All studies were completed in less than 45 minutes and required a mean of 12 breath holds (16 heartbeats). Of the total number of coronary segments, 91% (357/394) were visualized. Excellent or good image quality was achieved in 86% of the segments. Blinded analysis of the coronary arteries revealed sensitivity of 93% and specificity of 88% in the detection of coronary stenoses.

CONCLUSIONS:

The integrated environment of dRT provides a rapid and flexible scan protocol for MRCA while achieving wide anatomical coverage, high image quality, and reliable detection of coronary stenosis in short scan time.