Radiology. 2005 Feb;234(2):611-6. Epub 2004 Dec 15.

Detectability of catheters on bedside chest radiographs: comparison between liquid crystal display and high-resolution cathode-ray tube monitors.

Author information

1
Department of Radiology and Ludwig Boltzmann-Institute for Clinical and Experimental Radiologic Research, University of Vienna, Währinger Gürtel 18-20, A-1090 Vienna, Austria. martina.scharitzer@meduniwien.ac.at

Abstract

PURPOSE:

To compare observer performance with a flat-panel liquid crystal display (LCD) monitor and with a high-resolution gray-scale cathode-ray tube (CRT) monitor in the detection of simulated support catheters on bedside chest radiographs.

MATERIALS AND METHODS:

The ethics committee did not require approval or patient informed consent when this study began. Because of a change in regulations, before images were acquired the nature of the study and procedures were explained to patients or their relatives, and consent was then obtained. A total of 131 catheter fragments (12-14 per radiograph) were superimposed over 10 anteroposterior bedside chest radiographs obtained with storage phosphor technology. Images were displayed on an LCD monitor (1536 x 2048 matrix) and a CRT monitor (2048 x 2560 matrix). Five radiologists independently located the catheter fragments and rated their confidence in detection with bright and subdued ambient light. A two-way analysis of variance and the Friedman test were used for statistical analysis.

RESULTS:

There was no significant difference for either display type with respect to correctly detected catheter fragments (mean sensitivity, 56.6% and 56.0% for the CRT and the LCD monitors, respectively, with bright light and 61.2% for both monitors with subdued light). With both display types, detection rate with bright light decreased significantly (P < .05). False-positive rates and confidence ratings were not significantly affected by monitor type or ambient light.

CONCLUSION:

In a study with simulation of clinical conditions, performance of the LCD monitor and high-resolution CRT monitor for detection of support catheters on bedside chest radiographs was equivalent. With both displays, detection performance was equally reduced with bright ambient light.

PMID:
15601893
DOI:
10.1148/radiol.2342031297
[Indexed for MEDLINE]