Prediction of resorption rates for composite polylactide/hydroxylapatite internal fixation devices based on initial degradation profiles

J Biomed Mater Res B Appl Biomater. 2004 Oct 15;71(1):201-5. doi: 10.1002/jbm.b.30091.

Abstract

Mechanical and physicochemical outcomes were used to predict the resorption rate of polylactide-based internal fixation devices with and without incorporation of an osteoconductive buffer. Devices were degraded in vitro for 1 year. Addition of an osteoconductive buffer to a resorbable polymer internal fixation device controlled the rate of acid generation resulting from polymer hydrolysis. The pH of the physiological buffer remained neutral when it was exposed to the buffered screw, whereas the pH decreased to approximately 3.0 for the polylactide device. Neutralizaton of the acids generated during polymer hydrolysis increased the projected resorption time of the buffered device to 104 weeks in comparison to 71 weeks for a similar screw made from polylactide. In addition, the buffered device retained a higher percentage of its initial flexural strength throughout the course of degradation than the polylactide screw. The flexural strength of the polylactide screw decreased 20% during the first 4 weeks of polymer degradation, whereas the buffered device maintained its initial mechanical properties through 16 weeks of degradation.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Drug Stability
  • Durapatite / chemistry*
  • Hydrolysis
  • Internal Fixators*
  • Materials Testing
  • Polyesters / chemistry*
  • Prosthesis Failure