Format

Send to

Choose Destination
Opt Lett. 2004 Aug 1;29(15):1754-6.

Micromachined array tip for multifocus fiber-based optical coherence tomography.

Author information

1
Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 2M9, Canada. victor.yang@utoronto.ca

Abstract

High-resolution optical coherence tomography demands a large detector bandwidth and a high numerical aperture for real-time imaging, which is difficult to achieve over a large imaging depth. To resolve these conflicting requirements we propose a novel multifocus fiber-based optical coherence tomography system with a micromachined array tip. We demonstrate the fabrication of a prototype four-channel tip that maintains a 9-14-microm spot diameter with more than 500 microm of imaging depth. Images of a resolution target and a human tooth were obtained with this tip by use of a four-channel cascaded Michelson fiber-optic interferometer, scanned simultaneously at 8 kHz with geometric power distribution across the four channels.