J Pharm Biomed Anal. 2004 Sep 21;36(1):17-22.

A rapid and simple determination of A77 1726 in human serum by high-performance liquid chromatography and its application for optimization of leflunomide therapy.

Author information

1
Department of Clinical Pharmacy & Clinical Pharmacology, Medical Centre Leeuwarden, P.O. Box 888, 8901 BR Leeuwarden, The Netherlands. e.n.van.roon@znb.nl

Abstract

Leflunomide is a disease-modifying antirheumatic drug, which is bioactivated by formation of A77 1726. In this study a rapid and simple quantitative assay using a reversed phase HPLC-UV method is validated for detection of A77 1726 in human serum. The HPLC-UV method uses a mobile phase consisting of methanol and a KH2PO4-buffer (45 mM, pH = 3) (50:50,v/v), at a flow rate of 1 mL/min. A77 1726 is detected by UV-absorption at 295 nm with a retention time of 8.9 min. Demoxepam is used as internal standard. Validation showed lower and upper limits of quantitation of 0.5 and 100 mg/L, respectively. The assay was linear over the concentration range of 0.5-100 mg/L (r2 > 0.999). Intra- and inter-day precision showed coefficients of variation within 15% over the complete concentration range; accuracy was within 8%. Commonly prescribed drugs to treat rheumatoid arthritis like disease-modifying antirheumatic drugs, analgesics and corticosteroids, and their main metabolites, are separated from A77 1726 with a resolution >2. Serum levels of A77 1726 in 37 patients on leflunomide therapy were determined using this HPLC-UV method. Measured serum A77 1726 serum concentrations in patient samples showed large variability with a range of 3-176 mg/L.

PMID:
15351043
DOI:
10.1016/j.jpba.2004.05.019
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Elsevier Science
Loading ...
Support Center