Amino acid analysis using disposable copper nanoparticle plated electrodes

Analyst. 2004 Sep;129(9):841-5. doi: 10.1039/b401573h. Epub 2004 Jul 21.

Abstract

A disposable copper nanoparticle-plated screen-printed carbon electrode (designated as Cun-SPE(100-nm)) provides a new material for the determination of native amino acids. All 20 underivatized amino acids can be sensitively determined at 0.0 V vs. Ag/AgCl in pH 8 phosphate buffer solution. The precisely controlled copper nanoparticles can boost up the CuIIO/CuI2O redox signal on the working surface without any prior pretreatment procedure. The formation of a reversible 1:1 CuIIO-amino acid complex on the Cun-SPE(100-nm) was proposed to play a key role in the reaction mechanism. Stable detection responses were obtained for all amino acids by flow injection analysis with detection limits (S/N = 3) that lie in the range of 24 nM-2.7 microM. Selected amino acids from six representative chemical natures were separated by HPLC and detected at the Cun-SPE(100-nm) with promising results.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Amino Acids / analysis*
  • Copper
  • Electrochemistry / instrumentation
  • Electrochemistry / methods
  • Microelectrodes
  • Nanotechnology

Substances

  • Amino Acids
  • Copper