Magn Reson Med. 2004 Jun;51(6):1194-204.

Dynamic field map estimation using a spiral-in/spiral-out acquisition.

Author information

1
Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA. bsutton@uiuc.edu

Abstract

The long readout times of single-shot acquisitions and the high field strengths desired for functional MRI (fMRI) using blood oxygenation level-dependent (BOLD) contrast make functional scans sensitive to magnetic field inhomogeneity. If it is not corrected during image reconstruction, field inhomogeneity can cause geometric distortions in the images when Cartesian k-space trajectories are used or blurring with spiral acquisitions. Many traditional methods to correct for field inhomogeneity distortions rely on a static field map measured with the use of images that are themselves distorted. In this work, we employ a regularized least-squares approach to jointly estimate both the undistorted image and field map at each acquisition using a spiral-in/spiral-out pulse sequence. Simulation and phantom studies show that this method is accurate and stable over a time series. Human functional studies show that the jointly estimated field map may be more accurate than standard field map estimates in the presence of respiration-induced phase oscillations, leading to better detection of functional activation. The proposed method measures a dynamic field map that accurately tracks magnetic field drift and respiration-induced phase oscillations during the course of a functional study.

PMID:
15170840
DOI:
10.1002/mrm.20079
[Indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Wiley
    Loading ...
    Support Center