Catheter Cardiovasc Interv. 2004 Jun;62(2):234-6.

Complication with an embolic protection device during carotid angioplasty.

Author information

1
Department of Interventional Cardiology, University of Padova Medical School, Padova, Italy. pcardaioli@yahoo.it

Abstract

A case of carotid artery stenting with an AngioGuard cerebral protection device is described in a patient with tight internal carotid stenosis. Deployment of AngioGuard induced sudden severe vasospasm, with subsequent minor stroke. According to our observations, AngioGuard may cause a vasospastic response with adverse neurologic effects.

PMID:
15170718
DOI:
10.1002/ccd.20061
[Indexed for MEDLINE]