Format

Send to

Choose Destination
Allergy. 1992 Oct;47(5):463-6.

Clinical evaluation of CAP System and RAST in the measurement of specific IgE.

Author information

1
1st Department of Internal Medicine, University of Milan, Italy.

Abstract

We investigated the diagnostic value of a new in vitro test, Pharmacia CAP System (Pharmacia Diagnostics AB, Uppsala, Sweden), for the quantitative measurement of allergen-specific IgE antibodies by comparison with RAST in 2 groups of patients, 71 atopic and 48 non-atopic. In the last 20 years RAST has supplied a good diagnostic tool, but this test presents some problems, the main one being sensitivity. The new test has a solid phase able to bind even very small amounts of specific IgE and an anti-IgE tracer with very low cross-reactivity with other immunoglobulins, thus presenting more favourable conditions. From the analysis of our results, Pharmacia CAP System gave higher sensitivity (94% compared to 88% of RAST) with no loss of specificity (96% for both tests). The reliability of these results is ensured by the proper selection of patients who were all suffering from pollinosis and were clinically diagnosed as certainly hypersensitive to a single pollen. A positive trend was found between severity of asthma and levels of specific IgE for timothy. Pharmacia CAP System appears to identify a larger number of atopic patients than RAST.

PMID:
1485648
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Wiley
Loading ...