Format

Send to

Choose Destination
J Chromatogr. 1992 Dec 23;584(2):270-4.

Determination by high-performance liquid chromatography of hydroxyurea in human plasma.

Author information

1
Department of Biochemistry, Royal North Shore Hospital, St. Leonards, N.S.W., Australia.

Abstract

An assay using reversed-phase high-performance liquid chromatography with electrochemical detection was developed to measure hydroxyurea in plasma at concentrations suitable for pharmacokinetic studies. The sample preparation is simple, the analysis rapid and assays can be batched. The between-run precision is excellent (coefficient of variation = 2.8-4.5%) and the limit of detection is 0.02 mmol/l. Preliminary studies have shown that the method is suitable for pharmacokinetic studies.

PMID:
1484113
DOI:
10.1016/0378-4347(92)80586-f
[Indexed for MEDLINE]

Supplemental Content

Loading ...