Anal Chem. 2004 Feb 1;76(3):781-7.

Attenuated total reflectance FTIR detection and quantification of low concentrations of aqueous polyatomic anions.

Author information

1
Department of Chemistry, Colorado State University, Fort Collins, CO 80523, USA.

Abstract

Development of a new quantitative method for determining low concentrations of aqueous polyatomic anions using attenuated total reflectance (ATR) FTIR spectroscopy is described. Evaporated thin-film coatings of anion-selective tetraalkylated ferrocenium salts were applied to the surface of ATR crystals, which enabled anion detection limits to be lowered up to 23 000-fold below those achieved using the commercially available spectrometer with identical uncoated ATR crystals. Linear calibration curves based on d(absorbance)/dt, which is related to the rate of anion exchange in the thin film, were established in the 0.04-30 microM range. Limits of detection (10-min analyses) for perchlorate, chlorate, trifluoromethanesulfonate, perfluoro-n-butanesulfonate, perfluoro-n-octanesulfonate, tetrafluoroborate, hexafluorophosphate, and pinacolylmethylphosphonate in aqueous solution were 0.03, 0.2, 0.05, 0.07, 0.06, 0.06, 0.6, and 0.7 microM, respectively, using the thin-film coatings. This simple detection/quantification method afforded good reproducibility with relatively fast detection times.

PMID:
14750876
DOI:
10.1021/ac034915i
[Indexed for MEDLINE]