Ann Thorac Surg. 2003 Oct;76(4):1167-70.

Continuous axial-flow left ventricular assist device (Jarvik 2000) maintains kidney and liver perfusion for up to 6 months.

Author information

1
Texas Heart Institute, St. Luke's Episcopal Hospital, and The University of Texas Medical School, Texas, Houston 77030, USA. george.v.letsou@uth.tmc.edu

Abstract

BACKGROUND:

The Jarvik 2000 axial flow left ventricular assist device (LVAD), under development for the past decade, has the potential to support patients temporarily until cardiac transplantation or as a permanent circulatory support, without the size limitations of other implantable systems.

METHODS:

To assess its ability to perfuse the kidneys and liver, we monitored renal and hepatic function in 10 patients who received the Jarvik 2000 LVAD as a bridge to transplantation. Left ventricular assistance was maintained for up to 214 days (> 6 months), and renal and hepatic function were monitored at least weekly.

RESULTS:

Renal function before LVAD implantation in these patients was normal in 7 (creatinine, < 1.5) and moderately impaired in 3 (creatinine, 1.2 to 2.0). Hepatic function was normal in 7 patients before LVAD implantation (total bilirubin< 1.2; serum glutamic-oxaloacetic transaminase (SGOT), < 40; serum glutamic-pyruvic transaminase (SGPT), < 50) and normal at the time of transplantation in all 10 patients. Of the 3 patients with abnormal hepatic function before LVAD implantation, 1 patient had also had moderate renal dysfunction.

CONCLUSIONS:

Despite reduced pulsatility, the Jarvik 2000 LVAD improves or maintains excellent renal and hepatic function during periods of circulatory assistance in patients awaiting transplantation.