IEEE Trans Biomed Eng. 1992 Oct;39(10):1032-44.

Adaptive filter for event-related bioelectric signals using an impulse correlated reference input: comparison with signal averaging techniques.

Author information

1
Institut de Cibernètica, Universitat Politècnica de Catalunya-CSIC, Barclona, Spain.

Abstract

Many bioelectric signals result from the electrical response of physiological systems to an impulse that can be internal (ECG signals) or external (evoked potentials). In this paper an adaptive impulse correlated filter (AICF) for event-related signals that are time-locked to a stimulus is presented. This filter estimates the deterministic component of the signal and removes the noise uncorrelated with the stimulus, even if this noise is colored, as in the case of evoked potentials. The filter needs two inputs: the signal (primary input) and an impulse correlated with the deterministic component (reference input). We use the LMS algorithm to adjust the weights in the adaptive process. First, we show that the AICF is equivalent to exponentially weighted averaging (EWA) when using the LMS algorithm. A quantitative analysis of the signal-to-noise ratio improvement, convergence, and misadjustment error is presented. A comparison of the AICF with ensemble averaging (EA) and moving window averaging (MWA) techniques is also presented. The adaptive filter is applied to real high-resolution ECG signals and time-varying somatosensory evoked potentials.

PMID:
1452169
DOI:
10.1109/10.161335
[Indexed for MEDLINE]