EPR dosimetric properties of formates

Appl Radiat Isot. 2003 Aug-Sep;59(2-3):181-8. doi: 10.1016/s0969-8043(03)00165-9.

Abstract

As a part of a program to develop an electron paramagnetic resonance (EPR) dosimeter suited for clinical use (doses in the cGy range), polycrystalline samples of lithium formate monohydrate (HCO2Li.H2O), magnesium formate dihydrate (C2H2O4Mg.2H2O), and calcium formate (C2H2O4Ca) have been examined. L-Alanine was included for comparison and reference. Samples were irradiated with 60Co gamma-rays and 60-220 kV X-rays. The dosimeter response was assessed using the peak-to-peak amplitude of the first-derivative EPR spectrum. Dose-response curves for the 60Co gamma-irradiated samples were constructed, and the dependences of the response on the photon energy, microwave power, and modulation amplitude were studied. Stability of the irradiation products upon storage (signal fading) was also investigated. Lithium formate monohydrate is by far the best candidate of the tested formates, suitable for measuring doses down to approximately 0.1 Gy. Lithium formate monohydrate is more sensitive than alanine by a factor of 5.6-6.8 in the tested photon energy range, it exhibits no zero-dose signal and shows a linear dose response in the dose range from 0.2 to 1000 Gy. Its EPR signal was found unchanged in shape and intensity 1 week after irradiation to 10 Gy. Various less favorable properties rendered the other formates generally unsuitable, although calcium formate exhibits some interesting EPR dosimetric properties.

Publication types

  • Comparative Study
  • Evaluation Study
  • Validation Study

MeSH terms

  • Electron Spin Resonance Spectroscopy / instrumentation*
  • Electron Spin Resonance Spectroscopy / methods*
  • Formates / chemistry*
  • Formates / radiation effects*
  • Gamma Rays
  • Materials Testing
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Radiometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers*

Substances

  • Formates