Format

Send to

Choose Destination
Chest. 2003 Jun;123(6):1939-46.

A spirometry-based algorithm to direct lung function testing in the pulmonary function laboratory.

Author information

1
Ottawa Hospital Research Institute, University of Ottawa, Ottawa, ON, Canada.

Abstract

OBJECTIVE:

To design a spirometry-based algorithm to predict pulmonary restrictive impairment and reduce the number of patients undergoing unnecessary lung volume testing.

DESIGN:

Two prospective studies of 259 consecutive patients and 265 consecutive patients used to derive and validate the algorithm, respectively.

SETTING:

A pulmonary function laboratory of a tertiary care hospital.

PATIENTS:

Consecutive adults referred to the laboratory for lung volume measurements and spirometry.

MEASUREMENTS:

The sensitivity of the algorithm for predicting pulmonary restriction and the cost savings associated with its use.

RESULTS:

Total lung capacity correlated strongly with FVC (r = 0.66) and showed an inverse correlation with the FEV(1)/FVC ratio (r = - 0.41). According to the algorithm, only patients with an FVC < 85% of predicted and an FEV(1)/FVC ratio >or= 55% required lung volume measurements following spirometry. The algorithm had a high sensitivity for predicting restriction and a high negative predictive value (NPV) for excluding restriction (sensitivity, 96%; NPV, 98%). The diagnostic properties of the algorithm were reproducible in the validation study. Application of the algorithm would eliminate the need for lung volume testing in 48 to 49% of patients referred to the pulmonary function test (PFT) laboratory, reducing costs by 33%.

CONCLUSIONS:

A spirometry-based algorithm accurately excludes pulmonary restriction and reduces unnecessary lung volume testing in the PFT laboratory almost in half.

PMID:
12796171
DOI:
10.1378/chest.123.6.1939