Format

Send to

Choose Destination
See comment in PubMed Commons below
J Am Soc Mass Spectrom. 2003 May;14(5):492-500.

Effect of different solution flow rates on analyte ion signals in nano-ESI MS, or: when does ESI turn into nano-ESI?

Author information

  • 1Institute for Pharmaceutical Chemistry, Johann Wolfgang Goethe University, Frankfurt a.M., Germany. Andrea.Schmidt@iachem.de

Abstract

In nano-ESI MS, the qualitative and quantitative characteristics of mass spectra vary considerably upon the use of different spraying conditions, i.e., aperture of the spraying needle and the voltage applied. The major parameters affected by the aperture size is the liquid flow rate which determines the initial droplet size and the current emitted upon the spray process, as described by different models of the ESI process. In the present study, the effect of flow rate on ion signals was studied systematically using mixtures of compounds with different physicochemical properties (i.e., detergent/oligosaccharide and oligosaccharide/peptide). For these model systems, the functional dependence of certain analyte-ion ratios upon the flow rate can be correlated to changes in analyte partition during droplet fission prior to ion release. Analyte suppression is practically absent at minimal flow rates below 20 nL/min.

PMID:
12745218
DOI:
10.1016/S1044-0305(03)00128-4
[PubMed - indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Springer Icon for Elsevier Science
    Loading ...
    Support Center