A DICOM-RT-based toolbox for the evaluation and verification of radiotherapy plans

Phys Med Biol. 2002 Dec 7;47(23):4223-32. doi: 10.1088/0031-9155/47/23/308.

Abstract

The verification of radiotherapy plans is an essential step in the treatment planning process. This is especially important for highly conformal and IMRT plans which produce non-intuitive fluence maps and complex 3D dose distributions. In this work we present a DICOM (Digital Imaging and Communication in Medicine) based toolbox, developed for the evaluation and the verification of radiotherapy treatment plans. The toolbox offers the possibility of importing treatment plans generated with different calculation algorithms and/or different optimization engines and evaluating dose distributions on an independent platform. Furthermore the radiotherapy set-up can be exported to the BEAM Monte Carlo code system for dose verification. This can be done by simulating the irradiation of the patient CT dataset or the irradiation of a software-generated water phantom. We show the application of some of the functions implemented in this toolbox for the evaluation and verification of an IMRT treatment of the head and neck region.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Head and Neck Neoplasms / radiotherapy
  • Humans
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radiotherapy Planning, Computer-Assisted*
  • Radiotherapy, Conformal / methods
  • Software
  • Tomography, X-Ray Computed